To make Spray Monitor configuration faster and easier, you can set up one Spray Monitor (the source) and then copy its configuration settings (its recipe) to other Monitors (the destinations).
NOTE: This function is enabled only when One-Button Recipes are disabled.
Touch the information and setup bar on the source Spray Monitor faceplate.
Select Copy Spray Monitor Configuration from the drop-down menu. The Copy Spray Monitor Configuration dialog box appears.
The top field is a drop-down list of Spray Monitor groups. Select the group in which the destination monitors belong, or select ALL to list all Spray Monitors.
NOTE: The name of the Spray Monitor you are copying from (the source) appears in the middle field (with a black background). The lower field lists the Spray Monitors in the group you selected, or if you selected ALL, all Spray Monitors.
Select the Spray Monitors to which you want to copy the configuration settings:
If the destination Spray Monitors have the same transducer, CO-plate, nozzle flow rate, and calibration spray count as the source Spray Monitor, check the Copy Calibration Configuration box.
NOTE: If you check the Copy Calibration Configuration box, you may have to calibrate the destination Spray Monitors.
Touch the Group button. The settings are copied to the selected Spray Monitors.