Alarm: Short Trigger Duration

Trigger duration is shorter than the calibrated trigger value.

Possible Cause

Resolution

Timer problem

Verify the timer settings and operation. Check timer output stability. If using a PLC as a timer, check for scan time jitter.

Spray duration set shorter after calibration

Recalibrate the spray monitor to the new settings.

Electrical glitches and/or noise in trigger interface to timer

Check the wiring between the timer and the iTrax Spray Monitor. Verify cable shielding.